DISPLAY
Display LCD Touch Panel 6.5" x 5", function keys are displayed for the operation being performed Connectors RJ-12 for well cable, USB for printer
Power This system uses XP Power Supply for medical use, Model #PDM60US15
Line Voltage 100 to 240 VAC, auto selectable by the power supply, 1.5 – 0.75 amps
Line Frequency 50/60 Hz, detachable line cord, built-in EMI filter and transient suppression
Auxiliary Port RS-232 connector, used for data export and firmware updates
Memory Stores wipe, calibration, background, high voltage, isotope specification, isotope efficiency, Chi-Square testing results, technologist list, wipe locations list and latest MDA calculation. Results can be displayed and printed.
Preset Radionuclides 27 including Tc-99m, Co-57, Cs-137, Ga-67, Tl-201, I-123, I-125, I-131, In-111, F-18
DETECTOR CABLE
Length Standard 8 ft. (243 cm); custom cable lengths available up to 20 meters
Conductors Six total – 2 for power, two for chassis ground, two for serial data
Connectors RJ-12
|
|
WELL COUNTER
Detector 2" x 2" NaI (Tl) integral line scintillation detector with a 0.75" dia x 1.44 " depth well (1.9 x 3.7 cm)
Style Remote Detector
Channels 64 MCA: Integral to Well Counter Spectral
Resolution FWHM 10%
Count Rate (Maximum) 15,000 cps
Connectors RJ-12 for power and signal
Lead Shielding 0.5" (1.2 cm) integral lead shield
Optional Calibration Source 0.1 μCi Cs-137 Rod Source
Optional Additional Shield 0.5" (1.2 cm) lead shield, slides over integral shield for total of 1.0" of shielding – includes cover
Cover Fits optional additional lead shield, has 0.25" (.6 cm) lead shielding
PHYSICAL DATA
Display Unit
Size 9.5" w x 12" h x 12" d (24.1 x 30.5 x 30.5 cm)
Weight 4.2 lb (1.9 kg)
WELL COUNTER
Size 6" d x 11" h (15.24 x 27.9 cm)
Weight 29 lb (3.2 kg)
Approvals ETL listed to UL 60601-1 and CAN/CSA C22.2 No. 601-1M90, IEC 60601-1, IEC 60601-1-4 and IEC 60601-1-2 086-331
Warranty Two Year
|