  • Axial and lateral resolution targets at depths of 3, 8 and 14 cm for precise resolution measurements of any ultrasound system
  • Anechoic cysts of 2, 4 and 6 mm diameter
  • -6 dB, +6 dB and high scatter grey scale targets
  • Convertible water dam for gel or water coupled scanning
  • Integral cover to protect scanning surface
  • Durable ergonomic design for ease of handling
  • Available in 0.5 and 0.7 dB/cm/MHz attenuation

The Multi-Purpose Gray Scale Phantom 403GS LE offers new advanced technology for measuring the image quality of high resolution ultrasound systems. Gray scale targets are provided for monitoring contrast and temporal resolution, distinguishing different intensities of brightness and border delineation capabilities of the ultrasound system. The 403 LE has the same general target layout and specifications as the 403GS LE but does not include grey scale targets.

The Multi-Purpose Gray Scale Phantom offers carefully placed targets to measure resolution, depth of penetration and electronic caliper distance accuracy. Axial resolution resolution tests. The resolution patterns as well as all vertical and horizontal targets are constructed of 0.1 mm nylon fibers for better detail definition and spatial resolution. Grey scale targets are set at -6 dB, +6 dB with high scatter relative to the background material and with attenuation properties equivalent to the Tissue Mimicking (TM) material. A 10 mm anechoic cyst is also provided to evaluate system noise and geometric distortion.

The 403GS LE and 403 LE phantoms incorporate our new Tissue Mimicking (TM) gel which provides a smoother background texture than conventional tissue mimicking gels. This gel reduces the backscatter that is inherent in other TM gels and is fully compatible with the latest in tissue harmonics equipment and pin spacing patterns are small, offering better axial technology. The 403GS LE also has a composite film scanning surface that has improved transmission properties so more of the ultrasonic beam can be transmitted and received.

Optional accessories for the Multi-Purpose Gray Scale Phantom include Soft Foam-Lined Carrying Case, Rigid Case, or the Precision Ultrasound Transducer Guide.

403LE-0.5 Multi-Purpose Gray Scale Phantom, 0.5 attenuation

403GS LE-0.5 Multi-Purpose Gray Scale Phantom, 0.5 attenuation, includes gray scale targets

403GS LE-0.7 Multi-Purpose Gray Scale Phantom, 0.7 attenuation, includes gray scale targets

 

Ultrasound Multi-Purpose Gray Scale Phantom

Ultrasound Multi-Purpose Gray Scale Phantom

This ultrasound image demonstrates a sector scan of the Multi-Purpose Grey Scale Phantom 403GS LE

SPECIFICATIONS

Tissue Mimicking Material
Speed of sound 1540 ±10 m/s at 22°C
Attenuation 0.5 or 0.7 ±0.05 dB/cm/MHz

Targets Anechoic Cysts
Diameters
2, 4, 6 and 10 mm
Speed of sound 1540 ±10 m/s at 22°C
Attenuation 0.05 ±0.01 dB/cm/MHz

Grey Scale Targets
Contrast
-6, +6 and +12dB relative to background
Diameter 10 mm
Speed of sound 1540 ±10 m/s at 22°C same as background

Pin Targets
Diameter of nylon lines
0.1 mm (0.004 in)
Vertical spacing 20 mm at 2 to 16 cm deep
Horizontal spacing 30 mm at 2 and 12 cm deep
Axial resolution 3, 8 and 14 cm

Construction
Scanning surface
Composite Film
Walls Extruded ABS
Dimensions 23.2 x 8.25 x 18.5 cm (9.25 x 3.25 x 7.25 in)
Weight 2.8 kg (6 lbs 2.5 oz)

All acoustic measurements at 4.5 MHz, 22°C